AquaVENT® Heater Humidifier Setup
Switch on Heater Humidifier with button on the right side. Heater will carry out a short self check.
Heater will default to ‘Invasive Mode’. Green light will be visible beside the Invasive Mode icon.
The Heater Humidifier will lock automatically after 2 mins. To unlock and lock, press and hold the ‘Function’ button for 3 seconds. Confirmation on the screen will display LOCK or UNLK.
Switch between Invasive and Non-invasive modes by pressing the ‘Mode’ button when the device is unlocked.
To change Temperature display, press the ‘Function’ button when the device is unlocked. Select from Chamber Temperature,
Heater Plate Temperature (used for testing the device), or
Airway Temperature Mode.
When all settings have been set, lock the device my pressing and holding the ‘Function’ button again for 3 seconds.
Should an alarm sound, the issue will be highlighted by a red light beside the error description on the left of the device.
Press ‘Mute’ to temporarily silence the alarm while the error is resolved.